MU-Analog-65-Ex

/, ex-zone/MU-Analog-65-Ex

MU-Analog-65-Ex

Ringbalance as “Passive Sensor” according to EN50020 (5.4) intrinsically safe “i”.
to be mounted in the hazardous area, Ex-zone 2.

Categories: ,

Description

Ringbalance as “Passive Sensor” according to EN50020 (5.4) intrinsically safe “i”.
to be mounted in the hazardous area, Ex-zone 2.

1. Ringbalance MU-Analog-Ex

– Housing for wall mounting IP65; see page 9
– Display large display 150 x 150 mm (6″ x 6″)
– Measuring Ranges from 0…40 Pa (+/-20 Pa) to 0…1.8 kPa
– Accuracy max. error +/-1.5% of span, or +/-1.5 Pa
– Scale Units Pa, kPa, daPa, mbar, mmWS, mmCE, in.W.C.
– Ambient Temper. -20 . . . 0 . . . +40 C° (-4 . . . +104 °F)
– Measuring Pickup Magnetic field sensor; non-contact and with infinite resolution

2. Ex-Transmitter PR5114B2A

– converts the sensor signal of the Ringbalance into the standard output
– signals of 0-20mA, 4-20mA, 0-10V, and 4-20mA (loop-powered)
– Approvals ATEX 0539 “Ex” II (1) G; [EEx ia] II;
– DEMKO 99 (ATEX 124571) applicable zone 1,2
– Housing snap-in mounting on DIN-rail,
– 109mm high, 24mm wide, 130mm deep
– Power supply 22…250V AC / 20…300V DC
– Outputs 0/4-20 mA; 0-10 V; and 4-20 mA (loop-powered)
– linear/proportional for draft/pressure/diff.-press;
– or with root-extraction for velocity/flow

Additional information

Gehäuse IP-Schutzart

IP65